APPLICATION

The whole instrument consists of a closed test chamber, a photometer measurement system, a radiation cone, a combustion system, an igniter, a test box, a bracket, a temperature measuring instrument and a smoke density test software; the  circuit  is  developed  by  a  single-chip  microcomputer, with  high  technical  content and  stable performance. This instrument is suitable for all solid  materials, such as foam, rubber, textile coverings,  painted surfaces, wood and other materials, and is widely used by production plants in the foam industry, solid material industries, and scientific research and testing units.



MAIN TECHNICAL INDICATORS AND WORKING CONDITIONS:

Inner Box Size

L914*W610*H914mm

Equipment Size

L1110*W750*H2040mm

Smoke Density)

0-924 Six-speed automatic shift

Measuring Range

0.0001100%

Measurement Accuracy

±3%

Working Temperature

Room temperature40

Relative Humidity

85%

Working Environment

When the instrument is running, it should avoid direct sunlight and no forced air flow.

 

 

Measurement Software

It operates on the WINDOWS XP interface, and the

main data that can be tested include smoke optical density, specific optical density, mass optical density, mass loss, mass loss rate, average smoke density, etc. It also has operation interfaces such as calibration, editing, testing, and parameter adjustment.

Lab VIEW style.

Displays the transmittance versus time curve.

Sample Size

75×75mm

Light Source Color Temperature

2200K±100K

Speed Of Light Diameter 51mm

The computer automatically records the specific optical density according to the light change.

 

Filter

Amplifying filter, optical density 2.0; calibration filter,    optical density 3.0; compensation filter, optical density 0.5, 3 pieces in total.

 

Ring Heater

There is a ring heater in the lower aperture, so that the temperature on the light window is 50 C-55 C,

which can minimize the smoke concentration on the surface.

Heating Cone

Power 2600W

Temperature control range: 0- 1000 C, thermocouple temperature measurement, equipped with automatic  cold junction compensation.

 

Heat Flow Radiometer

The heat flow meter signal is collected and calculated by computer, equipped with water cooling pipeline.

 

Auxiliary Heater

With a power of 500W, the internal temperature of the test chamber can be stabilized at 40 C±10 C or

55 C±10 C.

 

Test Sample Weight Change

The balance weighs 2000g, the weighing accuracy is 0.1g, and the height of the weighing support rod is   adjustable.

 

Gas

For propane with a purity of more than 95%, the minimum pressure is 3.5kPa±kPa.

 

Automation

Computer automatic control, acquisition, recording and settlement of results

Size

1.22M×0.68M×2M



Weight

About 500Kg

Power Supply

220V, 50Hz, 4KW